Modbus
®
Mapping Assignments for Micro Motion
®
Transmitters 11
Modbus Mapping Assignments continued
3 0310 High-order word for binary volume total
Ö
3
3 0311 Low-order word for binary volume total
Ö
3
3 0312 Least significant word for binary volume total
Ö
3
3 0419 Bit #0 (E)EPROM checksum error, core processor
ÖÖÖ
Bit #1 RAM test error, core processor
ÖÖÖ
Bit #2 Real-time interrupt failure
ÖÖÖ
Bit #3 Sensor not vibrating
ÖÖÖ
Bit #4 Temperature sensor out of range
ÖÖÖ
Bit #5 Calibration failure
ÖÖÖ
Bit #6 Other failure occurred
ÖÖÖ
Bit #7 Transmitter initializing/warming up
ÖÖÖ
Bit #8 Primary variable out of limits
Ö
Bit #9 Secondary variable out of limits
Ö
Bit #10 Not used
Bit #11 Not used
Bit #12 Watchdog error
Ö
Bit #13 Cold start occurred
Ö
Bit #14 Transmitter configuration changed
Ö
Bit #15 Transmitter fault
ÖÖÖ
3 0420 Bit #0 Primary mA output saturated
ÖÖ
Bit #1 Secondary mA output saturated
1
Ö
Bit #2 Primary mA output fixed
ÖÖ
Bit #3 Secondary mA output fixed
1
Ö
Bit #4 Density overrange
ÖÖÖ
Bit #5 Drive overrange
ÖÖÖ
Bit #6 Not used
Bit #7 External input failure
ÖÖ
Bit #8 (E)EPROM checksum failure, core processor
ÖÖÖ
Bit #9 RAM diagnostic failure, core processor
ÖÖÖ
Bit #10 Sensor not vibrating
ÖÖÖ
Bit #11 Temperature sensor failure
ÖÖÖ
Bit #12 Input overrange
ÖÖÖ
Bit #13 Frequency/pulse output saturated
ÖÖ
Bit #14 Transmitter not configured
ÖÖÖ
Bit #15 Real-time interrupt failure
ÖÖÖ
3 0421 Bit #0 Burst mode enabled
ÖÖ
Bit #1 Power reset occurred
ÖÖÖ
Bit #2 Transmitter initializing/warming up
ÖÖÖ
Bit #3 Not used
Bit #4 Not used
Bit #5 Event 2 ON
ÖÖÖ
Bit #6 Event 1 ON
ÖÖÖ
Bit #7 Sensor/transmitter communication failure
ÖÖ
Bit #8 Calibration failure
ÖÖÖ
Bit #9 Zero value too low
ÖÖÖ
Bit #10 Zero value too high
ÖÖÖ
Bit #11 Zero too noisy
ÖÖÖ
Bit #12 Transmitter electronics failure
ÖÖÖ
Bit #13 Data loss possible
ÖÖÖ
Bit #14 Calibration in progress
ÖÖÖ
Bit #15 Slug flow
ÖÖÖ
Table 2-5. Input registers continued
Address Description MVDSolo
Series
1000
Series
2000 RFT9739