6 Modbus
®
Mapping Assignments for Micro Motion
®
Transmitters
Modbus Mapping Assignments continued
2 0187 0188 Volume flow minimum range
ÖÖÖÖ
2 0189 0190 Flow rate internal damping (seconds)
ÖÖÖÖ
2 0191 0192 Temperature internal damping (seconds)
ÖÖÖÖ
2 0193 0194 Density internal damping (seconds)
ÖÖÖÖ
2 0195 0196 Mass flow cutoff for frequency/pulse output and totalizers
ÖÖÖÖ
2 0197 0198 Volume flow cutoff for frequency/pulse output and totalizers
ÖÖÖÖ
2 0199 0200 Slug flow high-density limit (g/cc)
ÖÖÖÖ
2 0201 0202 Slug flow low-density limit (g/cc)
ÖÖÖÖ
2 0203 0204 Primary mA output present current (milliamps)
ÖÖÖ
2 0205 0206 Added damping on primary mA output (seconds)
ÖÖÖ
2 0207 0208 Flow cutoff for primary mA output
ÖÖÖ
2 0209 0210 Primary variable at 20 mA
ÖÖÖ
2 0211 0212 Primary variable at 0 mA or 4 mA
ÖÖÖ
2 0213 0214 Secondary mA output present current (milliamps)
1
ÖÖ
2 0215 0216 Added damping on secondary mA output (seconds)
1
ÖÖ
2 0217 0218 Flow cutoff for secondary mA output
1
ÖÖ
2 0219 0220 Secondary variable at 20 mA
Ö
1
Ö
2 0221 0222 Secondary variable at 0 mA or 4 mA
Ö
1
Ö
2 0223 0224 Frequency setpoint or number of pulses (Hz)
ÖÖÖ
2 0225 0226 Flow rate or total represented by frequency or number of pulses
ÖÖÖ
2 0227 0228 Frequency pulse width (seconds)
ÖÖÖ
2 0229 0230 Frequency/pulse output present frequency (Hz)
ÖÖÖ
2 0231 0232 Flowmeter zeroing standard deviation
Ö
2 0233 0234 Present flow signal offset at zero flow
ÖÖÖÖ
2 0235 0236 Flowmeter zeroing standard deviation limit
Ö
2 0237 0238 Special mass unit conversion factor
ÖÖÖÖ
2 0239 0240 Special volume unit conversion factor
ÖÖÖÖ
2 0241 0242 Event 1 setpoint
ÖÖÖÖ
2 0243 0244 Event 2 setpoint
ÖÖÖÖ
2 0245 0246 1 (E)EPROM checksum failure
ÖÖÖÖ
2 RAM diagnostic failure
ÖÖÖÖ
4 Sensor failure
ÖÖÖÖ
8 Temperature sensor failure
ÖÖÖÖ
16 Input overrange
ÖÖÖÖ
32 Frequency/pulse output saturated
ÖÖÖÖ
64 Transmitter not configured
ÖÖÖÖ
128 Real-time interrupt failure
ÖÖÖÖ
256 Primary mA output saturated
ÖÖÖ
512 Secondary mA output saturated
1
ÖÖ
1024 Primary mA output fixed
ÖÖÖ
2048 Secondary mA output fixed
1
ÖÖ
4096 Density overrange
ÖÖÖÖ
8192 Flowmeter zeroing failure
ÖÖÖÖ
16384 Zero value too low
ÖÖÖÖ
32768 Zero value too high
ÖÖÖÖ
65536 Transmitter electronics failure
ÖÖÖÖ
131072 Flowmeter zeroing in progress
ÖÖÖÖ
262144 Slug flow
ÖÖÖÖ
524288 Power reset occurred
ÖÖÖÖ
1048576 Transmitter configuration changed
ÖÖÖÖ
2097152 Transmitter initializing/warming up
ÖÖÖÖ
Table 2-4. Floating-point register pairs continued
Address Description MVDSolo
Series
1000
Series
2000 RFT9739