Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
6-8
Rosemount Radar Master > Tools > Simulation Mode
Figure 6-8. Simulation Mode
Also, or alternatively, activate the transmitter's loop test function. Output 4, 12,
and 20 mA and verify that the readings in the DCS match up.
Rosemount Radar Master > Output > Analog Out > Loop Test
Figure 6-9. Loop test for Analog
Out 1
Echo Curve Verification
1. Download and review the echo-curve
Figure 6-10. The Echo Curve
presents all visible echoes
Product
Surface
Surface Threshold / ATC
Hold Off
distance
Reference
Unknown
False Echo
Tank
Bottom
Vedere la pagina 83