TUpH Combination pH/ORP Sensors396P and 396PVP396P + 396PVP Sensors Instruction ManualLIQ_MAN_396P_396PVPAugust 2013
4 Description and SpecificationsIntegral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial / 396PVP: none - must use matingVP cablePreamplifier Opt
Description and Specifications 5The 396PVP Sensor has similar features to the 396P. However,the 396PVP is offered with the Variopol (VP) connector and
6 Description and SpecificationsConnector cable, VP8 (required for all first time installations of VP sensors)24281-00 15 ft. (4.6m) VP8 cable24281-01
Installation 7Section 2: Installation2.1 Unpacking and InspectionInspect the outside of the carton for any damage. If damage is detected, contact the
8 Installation2.2.1 Flow Through and Insertion Mounting 396P and 396PVP Sensors have a 1-inch MNPT process connection at the front of the s
Installation 9Figure 2-2. Flow-Through Tee with Adapter (PN 915240-xx*)++ '&((&%%)"&%!'( $
10 InstallationFigure 2-4. 396P with Insertion Mounting Adapter (PN 23242-02). Not for use with 396PVP. Mounting adapterallows for sensor removal with
Installation 11Figure 2-6. Submersion Installations 40396P04 A%'"#&*%)
12 InstallationFigure 2-7. Spray Wash Assembly PN 4091-00Section 2: Installation 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396P
Wiring 13Section 3: Wiring3.1 General Figures in this section provide the guidelines for wiring the 396P-01 sensor to variousAnalyzer/Transmitter ins
This page left blank intentionally
14 WiringFigure 3-1. Wiring for 396P-01 (Gray Cable)and 54e pH/ORPSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_3
Wiring 15396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-5. Wiring for 396PVP (Gray Cable)and 54e pH
16 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-8. Wiring for 396P-02 and 1055Figure 3-9. W
Wiring 17396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-11. Wiring for Dual 396P-02 and 1055Figure
18 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-13. Wiring for 396P-01 (Gray Cable) and1056
Wiring 19396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-17. Wiring for 396PVP (Gray Cable) and1056/
20 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-20. Wiring for 396P-01 (Gray Cable) and1057
Wiring 21396P + 396PVP Sensors Instruction Manual Section 3: 3WiringLIQ_MAN_396P_396PVP August 2013Figure 3-24. Wiring for 396PVP (Gray Cable) and1057
22 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-27. Wiring for 396P-01 (Blue Cable) and 106
Wiring 23396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-29. Wiring for 396P-02 (Blue Cable) and 106
Emerson Process Management designs, manufactures and tests its products to meet many nationaland international standards. Because these sensors are so
24 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-31. Wiring for 396PVP (Blue Cable) and 1066
Wiring 25396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-33. Wiring for 396P-01 (Gray Cable) and5081
26 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-37. Wiring for 396PVP (Gray Cable) and5081-
Wiring 27396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-41. Wiring for 396P-01 (Gray Cable) and 608
28 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-43. Wiring for 396P-02 (Gray Cable) and 608
Wiring 29396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-45. Wiring for 396PVP (Gray Cable) and 6081
30 WiringFigure 3-47. Wiring for 396P-01 (Gray Cable)and XmtFigure 3-48. Wiring for 396P-01 (Blue Cable)and XmtFigure 3-49. Wiring for 396P-02 (Gray C
Wiring 31396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-51. Wiring for 396PVP (Gray Cable)and XmtFi
32 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-55. Wiring Details for 396PVP or 396P-02-54
Wiring 33Figure 3-58. Wiring Details for 396PVP or 396P-02-54with Mating Variopol Cable for use with 1054A/B &2054Figure 3-59. Wiring Details for
This manual contains instructions for installation and operation of the 396P and 396PVP Sensors. Thefollowing list provides s concerning all revisions
34 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPThis page left blank intentionally
Section 4: Start-Up and Calibration4.1 396P and 396PVP pH Sensors4.1.1 Sensor preparationShake down the sensor to remove any air bubbles that may be
2. Measure and record the pH of the process solution sample with another temperature compen-sated, calibrated pH instrument. For best results, standar
Maintenance 37Section 5: Maintenance5.1 General InformationThe 396P and 396PVP Sensors require minimum maintenance. The sensor should be kept cleanan
38 MaintenanceNOTE: Erroneous pH results may result immedi ately after acid soak, due to reference junctionpotential build-up. Replace the sensor if c
Maintenance 39ammonium sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a satu-rated KCl/AgCl reference electrode and platinum
40 MaintenanceThis page left blank intentionallySection 5: Maintenance 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVP
Diagnostics and Troubleshooting 41Section 6: Diagnostics and Troubleshooting6.1 54e/56/1056/1057/1066/3081/4081/5081/XMT Diagnostics and Troubleshoot
42 Diagnostics and Troubleshooting6.2 Troubleshooting without Advanced DiagnosticsTable 7-2, below, lists common problems, causes and remedies typica
Diagnostics and Troubleshooting 43TABLE 6-3. Model 396P and 396PVP pH/ORP Replacement Parts and AccessoriesPN DESCRIPTION QUANTITY11275-01 Sensor
ContentsSection 1: Description and Specifications1.1 Features and Applications ...
44 Diagnostics and TroubleshootingSection 6: Diagnostics & Troubleshooting 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVP
Diagnostics and Troubleshooting 45WARRANTYSeller warrants that the firmware will execute the programming instructions provided by Seller, and that the
46This page left blank intentionallySection 6: Diagnostics & Troubleshooting 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396P
Return of Material 47Section 7: Return of Material7.1 GeneralTo expedite the repair and return of instruments, proper communication between thecustome
LIQ_MAN_396P_396PVPRev. BAugust 2013Credit Cards for U.S. Purchases Only.8Emerson Process Management2400 Barranca ParkwayIrvine, CA 92606 USATel: (94
This page left blank intentionallyTable of Contents 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPii Table of Contents
Section 1: Description and Specifications1.1 Features and ApplicationsTUpH™pH/ORP-sensors are now offered with SMART capabilities. SMART option becom
allows the process to flow by the glass electrode for accurate and reliable pH measurement. BothpH glass bulbs — the standard hemi or optional flat pH
Description and Specifications 3The 396P and 396PVP sensors are available in two configurations: the standard shrouded tip, inwhich the pH glass is co