Pagina 1 - 396P and 396PVP

TUpH Combination pH/ORP Sensors396P and 396PVP396P + 396PVP Sensors Instruction ManualLIQ_MAN_396P_396PVPAugust 2013

Pagina 2

4 Description and SpecificationsIntegral Cable 396P: Code 01 - 25 ft; Code 02 - 15 ft coaxial / 396PVP: none - must use matingVP cablePreamplifier Opt

Pagina 3 - Essential Instructions

Description and Specifications 5The 396PVP Sensor has similar features to the 396P. However,the 396PVP is offered with the Variopol (VP) connector and

Pagina 4 - About This Document

6 Description and SpecificationsConnector cable, VP8 (required for all first time installations of VP sensors)24281-00 15 ft. (4.6m) VP8 cable24281-01

Pagina 5 - Contents

Installation 7Section 2: Installation2.1 Unpacking and InspectionInspect the outside of the carton for any damage. If damage is detected, contact the

Pagina 6

8 Installation2.2.1 Flow Through and Insertion Mounting 396P and 396PVP Sensors have a 1-inch MNPT process connection at the front of the s

Pagina 7 - 1.1 Features and Applications

Installation 9Figure 2-2. Flow-Through Tee with Adapter (PN 915240-xx*)++ '&((&%%)"&%!'( $

Pagina 8

10 InstallationFigure 2-4. 396P with Insertion Mounting Adapter (PN 23242-02). Not for use with 396PVP. Mounting adapterallows for sensor removal with

Pagina 9 - 1.2 Specifications - General

Installation 11Figure 2-6. Submersion Installations 40396P04 A%'"#&*%)

Pagina 10 - 1.3 Ordering Information

12 InstallationFigure 2-7. Spray Wash Assembly PN 4091-00Section 2: Installation 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396P

Pagina 11

Wiring 13Section 3: Wiring3.1 General Figures in this section provide the guidelines for wiring the 396P-01 sensor to variousAnalyzer/Transmitter ins

Pagina 12 - Accessories

This page left blank intentionally

Pagina 13 - Section 2: Installation

14 WiringFigure 3-1. Wiring for 396P-01 (Gray Cable)and 54e pH/ORPSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_3

Pagina 14 - 2.2.2 Submersion Mounting

Wiring 15396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-5. Wiring for 396PVP (Gray Cable)and 54e pH

Pagina 15

16 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-8. Wiring for 396P-02 and 1055Figure 3-9. W

Pagina 16 - 40396P02 A

Wiring 17396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-11. Wiring for Dual 396P-02 and 1055Figure

Pagina 17 - 40396P03 A

18 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-13. Wiring for 396P-01 (Gray Cable) and1056

Pagina 18 - 12 Installation

Wiring 19396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-17. Wiring for 396PVP (Gray Cable) and1056/

Pagina 19 - Section 3: Wiring

20 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-20. Wiring for 396P-01 (Gray Cable) and1057

Pagina 20

Wiring 21396P + 396PVP Sensors Instruction Manual Section 3: 3WiringLIQ_MAN_396P_396PVP August 2013Figure 3-24. Wiring for 396PVP (Gray Cable) and1057

Pagina 21 - Wiring 15

22 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-27. Wiring for 396P-01 (Blue Cable) and 106

Pagina 22 - 16 Wiring

Wiring 23396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-29. Wiring for 396P-02 (Blue Cable) and 106

Pagina 23 - Wiring 17

Emerson Process Management designs, manufactures and tests its products to meet many nationaland international standards. Because these sensors are so

Pagina 24 - 18 Wiring

24 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-31. Wiring for 396PVP (Blue Cable) and 1066

Pagina 25 - Wiring 19

Wiring 25396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-33. Wiring for 396P-01 (Gray Cable) and5081

Pagina 26 - 20 Wiring

26 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-37. Wiring for 396PVP (Gray Cable) and5081-

Pagina 27 - Wiring 21

Wiring 27396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-41. Wiring for 396P-01 (Gray Cable) and 608

Pagina 28 - 22 Wiring

28 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-43. Wiring for 396P-02 (Gray Cable) and 608

Pagina 29 - Wiring 23

Wiring 29396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-45. Wiring for 396PVP (Gray Cable) and 6081

Pagina 30 - 24 Wiring

30 WiringFigure 3-47. Wiring for 396P-01 (Gray Cable)and XmtFigure 3-48. Wiring for 396P-01 (Blue Cable)and XmtFigure 3-49. Wiring for 396P-02 (Gray C

Pagina 31 - Wiring 25

Wiring 31396P + 396PVP Sensors Instruction Manual Section 3: WiringLIQ_MAN_396P_396PVP August 2013Figure 3-51. Wiring for 396PVP (Gray Cable)and XmtFi

Pagina 32 - 26 Wiring

32 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPFigure 3-55. Wiring Details for 396PVP or 396P-02-54

Pagina 33 - Wiring 27

Wiring 33Figure 3-58. Wiring Details for 396PVP or 396P-02-54with Mating Variopol Cable for use with 1054A/B &2054Figure 3-59. Wiring Details for

Pagina 34 - 28 Wiring

This manual contains instructions for installation and operation of the 396P and 396PVP Sensors. Thefollowing list provides s concerning all revisions

Pagina 35 - Wiring 29

34 WiringSection 3: Wiring 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPThis page left blank intentionally

Pagina 36 - 30 Wiring

Section 4: Start-Up and Calibration4.1 396P and 396PVP pH Sensors4.1.1 Sensor preparationShake down the sensor to remove any air bubbles that may be

Pagina 37 - Wiring 31

2. Measure and record the pH of the process solution sample with another temperature compen-sated, calibrated pH instrument. For best results, standar

Pagina 38 - 32 Wiring

Maintenance 37Section 5: Maintenance5.1 General InformationThe 396P and 396PVP Sensors require minimum maintenance. The sensor should be kept cleanan

Pagina 39 - Wiring 33

38 MaintenanceNOTE: Erroneous pH results may result immedi ately after acid soak, due to reference junctionpotential build-up. Replace the sensor if c

Pagina 40 - 34 Wiring

Maintenance 39ammonium sulfate) will produce a nominal ORP of 476 +20 mV at 25°C when used with a satu-rated KCl/AgCl reference electrode and platinum

Pagina 41 - 4.1.2 pH Calibration

40 MaintenanceThis page left blank intentionallySection 5: Maintenance 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVP

Pagina 42 - 4.2.2 ORP Calibration

Diagnostics and Troubleshooting 41Section 6: Diagnostics and Troubleshooting6.1 54e/56/1056/1057/1066/3081/4081/5081/XMT Diagnostics and Troubleshoot

Pagina 43 - Section 5: Maintenance

42 Diagnostics and Troubleshooting6.2 Troubleshooting without Advanced DiagnosticsTable 7-2, below, lists common problems, causes and remedies typica

Pagina 44 - 5.4 396P and 396PVP ORP

Diagnostics and Troubleshooting 43TABLE 6-3. Model 396P and 396PVP pH/ORP Replacement Parts and AccessoriesPN DESCRIPTION QUANTITY11275-01 Sensor

Pagina 45 - Maintenance 39

ContentsSection 1: Description and Specifications1.1 Features and Applications ...

Pagina 46 - 40 Maintenance

44 Diagnostics and TroubleshootingSection 6: Diagnostics & Troubleshooting 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVP

Pagina 47

Diagnostics and Troubleshooting 45WARRANTYSeller warrants that the firmware will execute the programming instructions provided by Seller, and that the

Pagina 48

46This page left blank intentionallySection 6: Diagnostics & Troubleshooting 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396P

Pagina 49

Return of Material 47Section 7: Return of Material7.1 GeneralTo expedite the repair and return of instruments, proper communication between thecustome

Pagina 50

LIQ_MAN_396P_396PVPRev. BAugust 2013Credit Cards for U.S. Purchases Only.8Emerson Process Management2400 Barranca ParkwayIrvine, CA 92606 USATel: (94

Pagina 51 - Irvine, CA 92606

This page left blank intentionallyTable of Contents 396P + 396PVP Sensors Instruction ManualAugust 2013 LIQ_MAN_396P_396PVPii Table of Contents

Pagina 52

Section 1: Description and Specifications1.1 Features and ApplicationsTUpH™pH/ORP-sensors are now offered with SMART capabilities. SMART option becom

Pagina 53 - Section 7: Return of Material

allows the process to flow by the glass electrode for accurate and reliable pH measurement. BothpH glass bulbs — the standard hemi or optional flat pH

Pagina 54

Description and Specifications 3The 396P and 396PVP sensors are available in two configurations: the standard shrouded tip, inwhich the pH glass is co