FCL with 1056 Analyzer
Instruction Manual
PN 51-FCL-1056 rev.E
March 2012
Vedere la pagina 0
1 2 3 4 5 6 ... 85 86

Sommario

Pagina 1 - FCL with 1056 Analyzer

FCL with 1056 AnalyzerInstruction ManualPN 51-FCL-1056 rev.EMarch 2012

Pagina 2 - ESSENTIAL INSTRUCTIONS

LIST OF FIGURESNumber Title Page2-1 Model FCL-01 ...

Pagina 4

1.3 SPECIFICATIONS — GENERALSample requirements:Pressure: 3 to 65 psig (122 to 549 kPa abs) A check valve in the inlet prevents the sensorflow cells

Pagina 5 - Temperature

31.6 ORDERING INFORMATIONFCL Free Chlorine Measuring System. The FCL is a complete system for the determination of free chlorine inaqueous samples. It

Pagina 6 - Diagnostics (on or off)

This page left blank intentionally4MODEL FCL-1056 SECTION 1.0DESCRIPTION AND SPECIFICATIONS

Pagina 7 - About This Document

5SECTION 2.0. INSTALLATIONMODEL FCL-1056 SECTION 2.0INSTALLATION2.1 UNPACKING AND INSPECTIONInspect the shipping container. If it is damaged, contact

Pagina 8 - FCL-1056

6MODEL FCL-1056

Pagina 9 - TABLE OF CONTENTS CONT’D

7FIGURE 2-1. Model FCL-01FIGURE 2-2. Model FCL-02INCHMILLIMETERINCHMILLIMETERMODEL FCL-1056

Pagina 10 - LIST OF TABLES CONT’D

This page left blank intentionally8MODEL FCL-1056

Pagina 11 - SECTION 1.0

SECTION 3.0. WIRINGMODEL FCL-1056

Pagina 12 - 1.5 SPECIFICATIONS — ANALYZER

ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!Your purchase from Rosemount Analytical, Inc. hasresulted in one of the finest instruments ava

Pagina 15 - INSTALLATION

This page left blank intentionally12MODEL FCL-1056

Pagina 17 - FIGURE 2-2. Model FCL-02

14MODEL FCL-1056 SECTION 4.0DISPLAY AND OPERATION4.2 KEYPADLocal communication with the analyzer is through the membrane keypad. Figures 4-4 and 4-5 e

Pagina 18

4.3 PROGRAMMING THE ANALYZER—TUTORIAL Setting up and calibrating the analyzer is easy. The following tutorial describes how to move around in theprogr

Pagina 19 - SECTION 3.0

164.4 SECURITY 4.4.1 How the Security Code WorksSecurity codes prevent accidental or unwanted changes to program settings or calibrations. There are t

Pagina 20 - 3.2 SENSOR WIRING

174.5 USING HOLD 4.5.1 PurposeTo prevent unwanted alarms and improper operation of control systems or dosing pumps, place the alarms andoutputs assign

Pagina 21

181. Press MENU. The main menu screen appears. Move the cursor toDisplay and press ENTER.2. The screen shows the present configuration. There are four

Pagina 22

19SECTION 5.0PROGRAMMING THE ANALYZER5.1 GENERALThis section describes how to make the following program settings using the local keypad. a. Configu

Pagina 23 - DISPLAY AND OPERATION

QUICK START GUIDEFOR FCL ANALYZER1. Refer to Section 2.0 for installation instructions and Section 3.0 for wiring instructions.2. Once connections

Pagina 24 - 4.2 KEYPAD

20MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZERTABLE 5-1. DEFAULT SETTINGSITEM CHOICES DEFAULTSensor assignment1. Sensor 1chlorine chlorine2.

Pagina 25 - MODEL FCL-1056 SECTION 4.0

21MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZERTABLE 5-1. DEFAULT SETTINGS (continued)ITEM CHOICES DEFAULTMeasurement1. Chlorine (sensor 1)a.

Pagina 26

22MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.3 CONFIGURING, RANGING, AND SIMULATING OUTPUTS.5.3.1 Purpose This section describes how to confi

Pagina 27

23MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.3.3. Procedure – Configure Outputs.1. Press MENU. The main menu screen appears. Move the cursor

Pagina 28

24MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.3.4. Procedure – Ranging Outputs.1. Press MENU. The main menu screen appears. Move the cursor to

Pagina 29 - PROGRAMMING THE ANALYZER

25MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.3.5 Procedure – Simulating Outputs1. Press MENU. The main menu screen appears. Move the cursor t

Pagina 30 - Sensor assignment

5.4.2 Definitions1. ASSIGNING ALARMS. There are four alarmsrelays. The relays are freely assignable to any sen-sor and to either the measurement (for

Pagina 31 - -999 to +999 mV 0 mV

27MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.4.3 Procedure – Configuring Alarms and Assigning Setpoints1. Press MENU. The main menu screen ap

Pagina 32 - MODEL FCL-1056 SECTION 5.0

28MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.4.4 Procedure – Simulating Alarms1. Press MENU. The main menu screen appears. Move the cursor to

Pagina 33

29MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.4.4 Procedure – Synchronizing Timers1. Synch Timers is available only if two or more alarm relay

Pagina 34

12. The main display appears. The outputs and alarms (if an alarm board is present) are assigned to default values.13. To change outputs, alarms, an

Pagina 35

30MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.5 CONFIGURING THE MEASUREMENT.5.5.1 PurposeThis section describes how to do the following:1. Pro

Pagina 36

MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.5.4 Procedure – Configuring the Measurement1. Press MENU. The main menu screen appears. Move the c

Pagina 37

MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.6 CONFIGURING TEMPERATURE RELATED SETTINGS5.6.1 PurposeThis section describes how to do the follow

Pagina 38

MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.7 CONFIGURING SECURITY SETTINGS5.7.1 PurposeThis section describes how to set security codes. Ther

Pagina 39

34MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.8 SETTING UP DIAGNOSTICS5.8.1 PurposeNOTEDiagnostic setup applies only to pH sensors. It appears

Pagina 40

35MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.8.3 Procedure – Setting Up Diagnostics1. Press MENU. The main menu screen appears. Move the curs

Pagina 41

36MODEL FCL-1056 SECTION 5.0PROGRAMMING THE ANALYZER5.9 RESETTING THE ANALYZER5.9.1 PurposeThis section describes how to clear user-entered values and

Pagina 42

MODEL FCL-1056 SECTION 6.0CALIBRATIONSECTION 6.0CALIBRATION6.1 INTRODUCTIONThe calibrate menu allows the user to do the following:1. Calibrate the t

Pagina 43

38MODEL FCL-1056 SECTION 6.0CALIBRATION3. Press MENU. The main menu screen appears. The cursor will be onCalibrate. Press ENTER.4. Choose the sensor y

Pagina 44

39MODEL FCL-1056 SECTION 6.0CALIBRATION6.3 CALIBRATION — FREE CHLORINE6.3.1 PurposeAs Figure 6-1 shows, a free chlorine sensor generatesa current dire

Pagina 45 - Diagnostic: On

MENU TREECalibrateSensor 1 (Free chlorine)ChlorineZeroIn processTemperatureSensor 2 (pH)pHBuffer CalAuto Select buffer (NIST, DIN19267, Ingold, Merck,

Pagina 46

40MODEL FCL-1056 SECTION 6.0CALIBRATION3. Choose the sensor you wish to calibrate. Sensor 1 is the chlorine sensor.Sensor 2 (if present) is the pH sen

Pagina 47 - CALIBRATION

MODEL FCL-1056 SECTION 6.0CALIBRATION6.3.3 Procedure-Calibrating the Sensor1. Place the chlorine sensor in the chlorine flow cell. If continuous (live

Pagina 48 - MODEL FCL-1056 SECTION 6.0

426.4 CALIBRATION – pH6.4.1 PurposeA pH sensor consists of a glass and reference electrode. Usually, the two electrodes are combined into a singlebody

Pagina 49

43MODEL FCL-1056 SECTION 6.0CALIBRATION3. Press MENU. The main menu screen appears. The cursor will be onCalibrate. Press ENTER.4. Choose the sensor y

Pagina 50

44MODEL FCL-1056 SECTION 6.0CALIBRATION8. Choose Start Auto Cal.If you wish to change the stability criteria or the pH buffer list from thedefault val

Pagina 51

45MODEL FCL-1056 SECTION 6.0CALIBRATION6.4.4 Procedure-Manual Calibration1. Obtain two buffer solutions. Ideally, the buffer pH values should bracke

Pagina 52 - The slope is reported as the

46MODEL FCL-1056 SECTION 6.0CALIBRATION10. Watch the pH reading for sensor 2 (S2) at the top of the screen. Oncethe reading is stable, enter the pH va

Pagina 53 - Buffer Cal

47MODEL FCL-1056 SECTION 6.0CALIBRATION6.4.5 Procedure-Standardization1. The pH value measured by the analyzer can be changed to match the reading fro

Pagina 54 - S2 pH Auto Cal

48MODEL FCL-1056 SECTION 6.0CALIBRATION6.4.6 Procedure-Entering a Known Slope and Offset1. Calibration data, i.e., slope and offset at pH 7, can be

Pagina 55

MODEL FCL-1056 SECTION 6.0CALIBRATION2. Press MENU. The main menu screen appears. The cursor will be onCalibrate. Press ENTER.3. Choose the output you

Pagina 56 - Manual Buffer 2

ProgramOutputsRange (assign values to 4 and 20 mA)ConfigureOutput 1 or 2Assign sensor and measurementRangeScaleDampeningFault mode (fixed or live)Faul

Pagina 57 - S1 Enter Value

50MODEL FCL-1056 SECTION 6.0CALIBRATIONThis page left blank intentionally

Pagina 58

51MODEL FCL-1056 SECTION 7.0DIGITAL COMMUNICATIONSSECTION 7.0DIGITAL COMMUNICATIONSTHE ANALYZER SUPPLIED WITH THE FCL DOES NOT HAVE THE DIGITAL COMMUN

Pagina 59

52MODEL FCL-1056 SECTION 7.0DIGITAL COMMUNICATIONSThis page left blank intentionally

Pagina 60

53MODEL FCL-1056 SECTION 8.0MAINTENANCESECTION 8.0MAINTENANCE8.1 ANALYZERThe analyzer used with the FCL needs little routine maintenance.Clean the ana

Pagina 61 - DIGITAL COMMUNICATIONS

MODEL FCL-1056 SECTION 8.0MAINTENANCE548.2.1 General.When used in clean water, the chlorine sensor requires little maintenance. Generally, the sensor

Pagina 62 - MODEL FCL-1056 SECTION 7.0

MODEL FCL-1056 SECTION 8.0MAINTENANCE5533523-00 Electrolyte Fill Plug 9550094 O-Ring, Viton 2-01433521-00 Membrane Retainer23501-08 Free Chlorine Memb

Pagina 63 - MAINTENANCE

56MODEL FCL-1056 SECTION 8.0MAINTENANCE8.4 CONSTANT HEAD FLOW CONTROLLER8.4.1 GeneralAfter a period of time, deposits may accumulate in the constant h

Pagina 67

About This DocumentThis manual contains instructions for installation and operation of the Model FCL-1056The following list provides notes concerning

Pagina 68

MODEL FCL-1056

Pagina 70

62MODEL FCL-1056

Pagina 71

63MODEL FCL-1056

Pagina 72

64MODEL FCL-1056

Pagina 73

MODEL FCL-1056 S

Pagina 74

66MODEL FCL-1056

Pagina 75

67MODEL FCL-1056

Pagina 76

68MODEL FCL-1056

Pagina 77

69MODEL FCL-1056

Pagina 78 - . The analyzer limits

iMODEL FCL-1056 TABLE OF CONTENTSFCL-1056 TABLE OF CONTENTSSection Title Page1.0 DESCRIPTION AND SPECIFICATIONS ...

Pagina 79

MODEL FCL-1056 SECTION

Pagina 80

9.9 SIMULATING INPUTS — pH9.9.1 GeneralThis section describes how to simulate a pH input intothe analyzer. To simulate a pH measurement, connect asta

Pagina 82

73NOTES

Pagina 83 - NOTES

A Worldwide Network of Sales and ServiceEmerson Process Management’s field sales offices are your source for more information on the fill line of Rose

Pagina 84

WARRANTYSeller warrants that the firmware will execute the programming instructions provided by Seller, and that the Goods manufacturedor Services pro

Pagina 85 - Irvine, CA 92606

Credit Cards for U.S. Purchases Only.The right people,the right answers,right now.ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITEhttp://www.rosemountan

Pagina 86 - The right people

Section Title Page7.0 DIGITAL COMMUNICATIONS .................... 518.0 MAINTENANCE ...