Reference Manual
00809-0100-4001, Rev JA
Section 4: Operation and Maintenance
November 2012
Operation and maintenance
Note
The range values for the 4 and 20 mA (1 and 5 Vdc) points should be at the nominal URV and
LRV. In the example above, the values are 1500 inH
2
O and 500 inH
2
O respectively. Confirm the
values on the HOME screen on the Field Communicator. Modify, if needed, by following the
steps in the Rerange section on page 53.