Series 3000 Transmitters and Discrete Controllers 7
Output signals
Non-intrinsically-safe output signals
Two active 4-20 mA outputs
Isolated to ± 50 VDC from all other outputs and earth
ground
Maximum load limit, 1000 ohms
Output is linear with process from 3,8 to 20,5 mA, per
NAMUR NE43 (June 1994)
Three discrete outputs
Configurable for the application
Power
Active or passive, software selectable
Voltage
24 VDC nominal
Current
Sourcing at 5,6 mA when V
out
= 3 VDC
Sinking up to 500 mA at 30 VDC maximum
One 2-wire frequency/pulse output
Configurable for rate control or totalization
Scalable to 10000 Hz
Output is linear with flow rate to 12500 Hz
Pulse width
50% duty cycle above 500 Hz
Configurable for frequencies below 500 Hz
Power
Active or passive, software selectable
Voltage
24 VDC nominal, active
20 VDC applied maximum, passive
Current
Sourcing 10 mA at 3 VDC, active
Sinking 500 mA, active or passive
Fault output
When a fault is detected, outputs go to configured states.
User can select high, low, last measured value, or internal
zero. Milliampere outputs conform to NAMUR NE43 (June
1994).
High output limits can be configured from 21 to 24 mA
for milliampere; 15000 Hz for frequency
Low output limits can be configured from 1 to 3,6 mA for
milliampere; 0 Hz for frequency
Last measured value is the last output for the process
variable
Internal zero drives the milliampere and frequency
outputs for process variables to zero values
Printer port
One RS-485 port (standard)
Requires external adaptor (not included) for conversion to
RS-232 signal
Send only
Power supply
AC power supply
Model 3300
85 to 265 VAC
50/60 Hz, 15 VA
0,25 A maximum at 85 VAC, 0,12 A maximum at 265 VAC
IEC 127-3/4 fuse, 0,63 A time-delay
Complies with low-voltage directive 73/23/EEC per IEC
1010-1 with Amendment 2
Installation (Overvoltage) Category II, Pollution Degree 2
Models 3350, 3500 and 3700
85 to 265 VAC
50/60 Hz, 30 VA
0,33 A maximum at 85 VAC, 0,15 A maximum at 265 VAC
IEC 127-3/4 fuse, 0,63 A time-delay
Complies with low-voltage directive 73/23/EEC per IEC
1010-1 with Amendment 2
Installation (Overvoltage) Category II, Pollution Degree 2
DC power supply
Model 3300
18 to 30 VDC
7 watts typical, 14 watts maximum
IEC 127-3/4 fuse, 1,6 A time-delay
Models 3350, 3500 and 3700
18 to 30 VDC
18 watts typical, 25 watts maximum
IEC 127-3/4 fuse, 1,6 A time-delay
Environmental limits
Ambient temperature limits
Operating –20 to 60°C
Storage –40 to 70°C
Some darkening of display might occur above 55°C
Display responsiveness decreases below –20°C
Humidity limits
5 to 95% relative humidity, non-condensing at 60°C
Vibration limits
Per IEC 68-2-6 at 1,0 g over 15 to 2000 Hz, 50 sweep
cycles
Environmental effects
EMI effects
All Series 3000 devices conform to NAMUR NE21
(June 1997)
Models 3300 and 3350
Meets EMC directive 89/336/EEC per EN 50081-2
(August 1993) and EN 50082-2 (March 1995)
Models 3500 and 3700
Meets EMC directive 89/336/EEC per EN 61326 Industrial
Ambient temperature effect
On analog outputs ± 0,005% of span per °C