MODEL XMT-C SECTION 9.0
PROGRAMMING THE TRANSMITTER
9.7 MAKING HART RELATED SETTINGS
For more information refer to Section 6.0.
9.8 RESETTING FACTORY CALIBRATION AND FACTORY DEFAULT SETTINGS
9.8.1 Purpose
This section describes how to install factory calibration and default values. The process also clears all fault mes-
sages and returns the display to the first quick start screen.
9.8.2 Procedure: Installing default settings
Outputs Temp
Measurement
>>>>
Security HART
RReesseett AAnnaallyyzzeerr
>>
Calibrate Hold
PPrrooggrraamm
Display
Load factory
settings?
YYeess
No
1. Press MENU. The menu screen appears. Choose Program.
2. Choose >>.
3. Choose ResetTransmitter.
4. Choose Yes or No. Choosing Yes clears previous settings and calibra-
tions and returns the transmitter to the first quick start screen.