Transmitter Configuration, Characterization, and Calibration
5.5.3 Temperature calibration
Temperature calibration is a two-point procedure: temperature offset calibration and temperature
slope calibration. The entire procedure must be completed without interruption. To perform a
temperature calibration, see Figure 5-7.
Figure 5-7 Temperature calibration
Fill sensor with
low-temperature fluid
Fill sensor with
high-temperature fluid
Calibration
Temp slope cal
Do Cal
Enter temperature of
high-temperature fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Wait until sensor achieves
thermal equilibrium
Calibration
Temp offset cal
Do Cal
Enter temperature of
low-temperature fluid
Calibration in Progress
light turns red
ProLink
Calibration in Progress
light turns green
Wait until sensor achieves
thermal equilibrium
Close Close