Emerson MicroCEM Specifiche Pagina 100

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 119
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 99
MAINTENANCE and SERVICE
can be disabled at slow sample rates, so there is no lag time between sampling and data
availability. The 16-bit interface further reduces software overhead by enabling all 16
A/D bits to be read in a single instruction, instead of requiring 2 8-bit read operations.
The net result of this streamlined design is that the ADIO board supports gap-free A/D
sampling at rates up to 200,000 samples per second, twice as fast as our previous
boards.
5.6.2.7 Specifications
Analog Inputs
Number of inputs
32 single-ended, 16 differential, or
16 SE + 8 DI; user selectable
A/D resolution 16 bits (1/65,536 of full scale)
Bipolar ranges ±10V, ±5V, ±2.5V, ±1.25V, ±0.625V
Unipolar ranges
0-10V, 0-5V, 0-2.5V, 0-1.25V, 0-
.625V,
Input bias current 100pA max
Overvoltage
protection
±35V on any analog input without
damage
Nonlinearity ±3LSB, no missing codes
Conversion rate 200,000 samples/sec.max
On-board FIFO 1K x 8(512 16-bit samples)
Calibration Automatic;values stored in EEPROM
Analog Outputs
Number of outputs 4
D/A resolution 12 bits (1/4096 of full scale)
Output ranges ±5, ±10, 0-5, 0-10
Output current ±5mA max per channel
Settling time 6µS max to 0.01%
Relative accuracy ±1 LSB
Nonlinearity ±1 LSB, monotonic
Reset All channels reset to OV
Calibration
Automatic; values stored in
EEPROM
Digital I/O
Main I/O 24 programmable I/O
Input current ±1µA max
Output current
Logic 0
64mA max per line
Logic 1
-15mA max per line
Auxilary I/O 4 inputs, 4 outputs, optional use as
Rosemount Analytical µCEM Continuous Analyzer Transmitter 5-12
Vedere la pagina 99
1 2 ... 95 96 97 98 99 100 101 102 103 104 105 ... 118 119